Quantum Cascade Laser-Based Vibrational Circular Dichroism Augmented by a Balanced Detection Scheme.

Daniel-Ralph HermannGeorg RamerMarkus Kitzler-ZeilerBernhard Lendl
Published in: Analytical chemistry (2022)
Vibrational circular dichroism (VCD) constitutes a powerful technique, enabling the determination of the absolute configuration of molecules without the need for specialized reagents. While delivering critical information, VCD signals commonly are several orders of magnitude weaker than classical absorbance signals, which so far necessitated long measurement times to achieve acceptable signal-to-noise ratios (SNRs) in VCD experiments. We present here an improved setup for the measurement of VCD in the range between 5.6 and 6.5 μm. Employing an external cavity quantum cascade laser (EC-QCL) as a high-power light source, we collected spectra with competitive noise levels in less than 5 min. The basis for this improvement was a balanced detection module combined with an optical path catered to VCD measurements. With the stabilization provided by the two-detector setup, noise originating from the laser source could be suppressed effectively. Noise level improvement up to a factor of 4 compared to the classical single detector EC-QCL-VCD could be reported. Compared to commercial Fourier transform infrared (FT-IR) instruments, the presented setup offers measurement time reductions of a factor of at least 6, with comparable noise levels. The applicability of the setup for qualitative and quantitative VCDs was proven. With the comparatively high temporal resolution provided, the monitoring of optically active processes will be possible in future applications.