Comparison of electrophysiological parameters between perimodiolar and lateral wall electrodes in paediatric cochlear implant users.

Andrzej ZarowskiAndrzej MoliszEmmanuel A M MylanusJ H M MiserusM LeblansJoost Js van DintherJanusz SiebertErwin F Offeciers
Published in: European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery (2020)
There exist significant differences in the contact impedance between different cochlear implant electrode arrays. These differences could influence the implant programming parameters and should be considered in any protocols related to automatic implant programming based on objective measures. The predictive value of the electrode impedances for the behavioural T-/C-levels is higher for the perimodiolar CA electrodes than for the lateral wall SS electrodes.