Combination of the best pacing configuration and atrioventricular and interventricular delays optimization in cardiac resynchronization therapy.
Pierre SocieFabien SquaraMarc SemichonOlivier ThomasAlain KhemachePhilippe RicciniPierre SquaraVincent AlgalarrondoGhassan MoubarakPublished in: Pacing and clinical electrophysiology : PACE (2018)
The combination of choosing the best quadripolar pacing configuration and optimizing atrioventricular and interventricular delays resulted in an improvement of cardiac output compared to standard biventricular stimulation in 68% of patients. The final cardiac output was comparable to multipoint pacing.