Login / Signup

Accuracy of low-weight versus standard syringe infusion pump devices depending on altitude.

Marc BlancherMaxence RepellinMaxime MaignanCyrielle ClapéArnaud PerrinJosé LabarèreGuillaume DebatyDamien Viglino
Published in: Scandinavian journal of trauma, resuscitation and emergency medicine (2019)
Despite satisfactory normative tests, low-weight SIPs deliver discontinuous flow with potential clinical implications for critically ill patients receiving vasoactive drugs. This study also highlights a thus far unknown negative impact of altitude on SIP function. We believe that normative requirements for SIP approval should be revised accordingly.
Keyphrases
  • body mass index
  • weight loss
  • physical activity
  • weight gain
  • low dose
  • body weight
  • risk assessment
  • human health