Optimized method for normal range estimation of standardized uptake values (SUVmax, SUVmean) in liver SPECT/CT images with somatostatin analog [99mTc]-HYNIC-TOC (Tektrotyd).

Hanna Piwowarska-BilskaHanna Piwowarska-BilskaBozena Birkenfeld
Published in: Nuclear medicine review. Central & Eastern Europe (2022)
Segmentation in the mid-coronal computed tomography (CT) image, at one-fourth of the height of the liver measured from the top, with a medium-sized volume of interest (VOI) outlined on a given transverse SPECT slice was regarded as the optimal method for estimating normal ranges for standardized uptake values. It is necessary to standardize quantification methods in the SPECT/CT studies. Our work is a step forward in obtaining standardization of SPECT/CT SUV calculation methods. Calculations for radiopharmaceutical uptake in tumors with volumes smaller than 5 mL are biased with a significant measurement error.