Login / Signup

Cardiac troponin measurement at the point of care: educational recommendations on analytical and clinical aspects by the IFCC Committee on Clinical Applications of Cardiac Bio-Markers (IFCC C-CB).

Paul CollinsonKristin Moberg AakreAmy SaengerRick BodyOle HammarstenAllan S JaffePete KavsakTorbjørn OmlandJordi Ordonez-LianosBrad KaronFred S Apple
Published in: Clinical chemistry and laboratory medicine (2023)
The International Federation of Clinical Chemistry and Laboarator Medicine (IFCC) Committee on Clinical Applications of Cardiac Bio-Markers (C-CB) has provided evidence-based educational resources to aid and improve the understanding of important analytical and clinical aspects of cardiac biomarkers. The present IFCC C-CB educational report focuses on recommendations for appropriate use, analytical performance, and gaps in clinical studies related to the use of cardiac troponin (cTn) by point of care (POC) measurement, often referred to as a point of care testing (POCT). The use of high-sensitivity (hs)-cTn POC devices in accelerated diagnostic protocols used in emergency departments or outpatient clinics investigating acute coronary syndrome has the potential for improved efficacy, reduction of length of stay and reduced costs in the health care system. POCT workflow integration includes location of the instrument, assignment of collection and testing responsibility to (non-lab) staff, instrument maintenance, in-service and recurrent training, quality control, proficiency assessments, discrepant result trapping, and troubleshooting and inventory management.
Keyphrases