Login / Signup

Design and validation of an open-source library of dynamic reference frames for research and education in optical tracking.

Alisa BrownAli UneriTharindu De SilvaAmir ManbachiJeffrey H Siewerdsen
Published in: Journal of medical imaging (Bellingham, Wash.) (2018)
Dynamic reference frames (DRFs) are a common component of modern surgical tracking systems; however, the limited number of commercially available DRFs poses a constraint in developing systems, especially for research and education. This work presents the design and validation of a large, open-source library of DRFs compatible with passive, single-face tracking systems, such as Polaris stereoscopic infrared trackers (NDI, Waterloo, Ontario). An algorithm was developed to create new DRF designs consistent with intra- and intertool design constraints and convert to computer-aided design (CAD) files suitable for three-dimensional printing. A library of 10 such groups, each with 6 to 10 DRFs, was produced and tracking performance was validated in comparison to a standard commercially available reference, including pivot calibration, fiducial registration error (FRE), and target registration error (TRE). Pivot tests showed calibration error [Formula: see text], indistinguishable from the reference. FRE was [Formula: see text], and TRE in a CT head phantom was [Formula: see text], both equivalent to the reference. The library of DRFs offers a useful resource for surgical navigation research and could be extended to other tracking systems and alternative design constraints.
Keyphrases