Effect of decentration, tilt and rotation on the optical quality of various toric intraocular lens designs: a numerical and experimental study.
Jesús Pérez-GraciaJorge AresFrancisco J ÁvilaLaura RemónPublished in: Biomedical optics express (2022)
Toric intraocular lenses (T-IOLs) may lose their optical quality if they are not correctly positioned inside the capsular bag once implanted. In this work, T-IOLs with cylinder powers of +1.50, +4.50 and +7.50 D and differing degrees of spherical aberration have been designed, manufactured and tested in vitro using a commercial optical bench that complies with the requirements of standard ISO 11979-2. Moreover, the effect of tilt and rotation on optical quality was assessed by means of numerical ray tracing on an astigmatic eye model, while the effect of decentration was evaluated numerically and experimentally.