Login / Signup

Benefit of the Use of GCxGC/MS Profiles for 1D GC/MS Data Treatment Illustrated by the Analysis of Pyrolysis Products from East Asian Handmade Papers.

Bin HanSilvia LobMichel Sablier
Published in: Journal of the American Society for Mass Spectrometry (2018)
In this study, we report the use of pyrolysis-GCxGC/MS profiles for an optimized treatment of data issued from pyrolysis-GC/MS combined with the automatic deconvolution software Automated Mass Spectral Deconvolution and Identification System (AMDIS). The method was illustrated by the characterization of marker compounds of East Asian handmade papers through the examination of pyrolysis-GCxGC/MS data to get information which was used for manually identifying low concentrated and co-eluting compounds in 1D GC/MS data. The results showed that the merits of a higher separation power for co-eluting compounds and a better sensitivity for low concentration compounds offered by a GCxGC system can be used effectively for AMDIS 1D GC/MS data treatment: (i) the compound distribution in pyrolysis-GCxGC/MS profiles can be used as "peak finder" for manual check of low concentration and co-eluting compound identification in 1D GC/MS data, and (ii) pyrolysis-GCxGC/MS profiles can provide better quality mass spectra with observed higher match factors in the AMDIS automatic match process. The combination of 2D profile with AMDIS was shown to contribute efficiently to a better characterization of compound profiles in the chromatograms obtained by 1D analysis in focusing on the mass spectral identification. Graphical abstract.
Keyphrases