Login / Signup

Accuracy of volume and pressure delivery by mechanical ventilators in use in neonatal intensive care units: A quality control study.

Valeria OttavianiChiara VeneroniValentina Dell'OrtoAnna LavizzariMaria Luisa VenturaMarta TossiciIwona MaroszynskaFabio MoscaPaolo TagliabueDaniele De LucaRaffaele L Dellaca
Published in: Pediatric pulmonology (2020)
Our findings suggest that loss of accuracy in ventilation variables is likely related to daily use of the devices rather than weakness in the design or manufacturing process, urging the improvement of maintenance and quality control procedures to preserve the performances of neonatal MVs during their entire lifespan.
Keyphrases
  • quality control
  • intensive care unit
  • mechanical ventilation
  • physical activity