A 10 J shock impedance in sinus rhythm correlates with a 65 J defibrillation impedance during subcutaneous defibrillator implantation using an intermuscular technique.
Toshimasa OkabeSalvatore J SavonaFaisal MattoChad C WardPrabhpreet SinghMuhammad Rizwan AfzalSteven J KalbfleischRaul WeissMahmoud HoumsseRalph S AugostiniJohn D HummelEmile G DaoudPublished in: Journal of cardiovascular electrophysiology (2021)
We demonstrated that 10 J shock impedance in SR correlated well with 65 J defibrillation impedance during IM S-ICD implantation. An IM implantation technique provides excellent generator location on postimplant CXR. The IM technique combined with 10 J shock in SR may be sufficient to predict and ensure the defibrillation efficacy of the S-ICD.