Login / Signup

Radiofrequency Ablation by a 21-Gauge Internally Cooled Electrode: Ex Vivo and In Vivo Evaluation by Rat Liver.

Haruyuki TakakiYuki KodamaHisao MiyamotoYuto IijimaYoshinari KikuchiHiroshi KodamaEisuke UeshimaKeiji NakashoReona WadaKoichiro Yamakado
Published in: Cardiovascular and interventional radiology (2020)
RF ablation using a 21-gauge electrode produced coagulation necrosis in the rat liver. The ablation volume became maximum when RF power of 7 W was applied with chilled water circulation.
Keyphrases
  • radiofrequency ablation
  • ultrasound guided
  • oxidative stress
  • carbon nanotubes
  • solid state
  • catheter ablation
  • atrial fibrillation