Evaluation of cutout factors with small and narrow fields using various dosimetry detectors in electron beam keloid radiotherapy.

Yu-Fang LinChen-Hsi HsiehHui-Ju TienYi-Huan LeeYi-Chun ChenLu-Han LaiShih-Ming HsuPei-Wei Shueng
Published in: Journal of X-ray science and technology (2024)
The parallel plate, semiflex chamber and EBT3 film are suitable dosimeters that are comparable with pinpoint 3D chambers in small and narrow electron fields. Notably, a semiflex chamber could be an alternative option to a pinpoint 3D chamber for cutout widths≥3 cm. It is very important to perform patient-specific cutout factor calibration with an appropriate dosimeter for keloid radiotherapy.