Accuracy of the end-expiratory lung volume measured by the modified nitrogen washout/washin technique: a bench study.
Joana Berger-EstilitaMatthias HaenggiDaniel OttDavid BergerPublished in: Journal of translational medicine (2021)
We conclude that the TestChest® simulator is an accurate simulation tool, concerning the simulation of lung volumes. The nitrogen wash-in/wash out method correlated positively with FRC changes, despite a relatively large bias in absolute measurements. The reference volumes in the lung simulator verified with CT volumetry were very close to their expected values. The reason for the bias could not be determined.