Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.
Yang ZhouYimei LiuMeining ChenJianlan FangLiangjie XiaoShaomin HuangZhenyu QiXiaowu DengJun ZhangYinglin PengPublished in: Journal of applied clinical medical physics (2024)
Dosimetry characteristics of the myQA SRS device meets the accuracy and sensitivity requirement of PSQA for SRS/SBRT treatment. The dose rate dependence should be adequately calibrated before its application and a more stringent GPR (3%/1 mm) evaluation criterion is suggested when it is used for SRS/SBRT QA.