Risk factors for distal migration of biliary plastic stents and related duodenal injury.

Xiang-Lei YuanLian-Song YeQin LiuChun-Cheng WuWei LiuXian-Hui ZengYu-Hang ZhangLin-Jie GuoYu-Yan ZhangYan LiXin-Yue ZhouBing Hu
Published in: Surgical endoscopy (2019)
The risk factors of distal migration are BBS and the length of the stent above the proximal end of the stricture. The risk factor of duodenal injury due to distal migration is large angle (≥ 30°) between the distal end of the stent and the centerline of the patient's body. Distal migration and related duodenal injury are more likely to present during the early period after biliary stenting.
Keyphrases
  • risk factors
  • minimally invasive