Three-electrode cell calorimeter for electrical double layer capacitors.

Joren E VosHendrik P RodenburgDanny Inder MaurTies J W BakkerHenkjan SiekmanBen H Erné
Published in: The Review of scientific instruments (2023)
A calorimeter was built to measure the heat from a porous capacitive working electrode connected in a three-electrode configuration. This makes it possible to detect differences between cathodic and anodic heat production. The electrochemical cell contains a large electrolyte solution reservoir, ensuring a constant concentration of the salt solution probed by the reference electrode via a Luggin tube. A heat flux sensor is used to detect the heat, and its calibration as a gauge of the total amount of heat produced by the electrode is done based on the net electrical work performed on the working electrode during a full charging-discharging cycle. In principle, from the measured heat and the electrical work, the change in the internal energy of the working electrode can be determined as a function of the applied potential. Such measurements inform about the potential energy and average electric potential of ions inside the pores, giving insight into the electrical double layer inside electrode micropores. Example measurements of the heat are shown for porous carbon electrodes in an aqueous salt solution.