An interventional MRI guidewire combining profile and tip conspicuity for catheterization at 0.55T.

Dursun Korel YildirimDogangun UzunChristopher G BruceJaffar M KhanToby RogersWilliam H SchenkeRajiv RamasawmyAdrienne E Campbell-WashburnDaniel A HerzkaRobert J LedermanOzgur Kocaturk
Published in: Magnetic resonance in medicine (2022)
The active guidewire was clearly visible via real-time MRI at 0.55T and exhibits a favorable geometric sensitivity profile depicting the guidewire continuously from shaft-to-tip including a unique curved-tip signature. RF-induced heating is clinically acceptable. This design allows safe device navigation through luminal structures and heart chambers. The detachable connector allows delivery and exchange of cardiovascular catheters while maintaining guidewire position. This enhanced guidewire design affords the expected performance of X-ray guidewires during human MRI catheterization.