SIZE-SPECIFIC DOSE ESTIMATES IN FETAL COMPUTED TOMOGRAPHY.

Masanao KobayashiYusei NishiharaTomonobu HabaYuta MatsunagaKazuyuki MinamiYasuki Asada
Published in: Radiation protection dosimetry (2022)
During fetal computed tomography (CT) imaging, because of differences in the pregnancy period and scanning conditions, different doses of radiation are absorbed by the fetus. We propose a correction coefficient for determining the fetal size-specific dose estimate (SSDE) from the CT dose index (CTDI) displayed on the console at tube voltages of 80-135 kVp. The CTDIs corresponding to pregnant women and fetuses were evaluated using a Monte Carlo (MC) simulation, and the ratio of these CTDIs was defined as the Fetus-factor. When the effective diameter of a fetus was approximately 10 cm, the Fetus-factor was 1.0. The estimated pregnant SSDE was multiplied by the Fetus-factor to estimate the fetal SSDE, which was compared with the fetal dose obtained by the MC simulation of the image of the fetal CT examination. The fetal dose could be estimated with an error of 31.5% in fetal examinations conducted using helical CT.