Effect of inter-electrode RF coupling on heating patterns of wire-like conducting implants in MRI.

Bhumi BhusalPallab BhattacharyyaTanvir BaigStephen JonesMichael Martens
Published in: Magnetic resonance in medicine (2022)
The level of RF-induced heating during an MR scan is affected significantly by RF coupling when more than one wire-like implant is present. In some of the configurations studied, the heating was increased by the presence of a second conductor partially immersed in the phantom. Thus, RF coupling is an important factor to consider in the assessment of safety issues for MRI when multiple implants are present.