A device for measuring sternal bone connectivity using vibration analysis techniques.

Atte JoutsenJuha HautalahtiEsa JaatinenSirkka GoebelerAntti PaldaniusJari ViikJari LaurikkaJari Hyttinen
Published in: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine (2019)
Technical solutions and description of validation process were given. The device was calibrated, and the vibration transmittance analysis differentiated intact and cut polylactide rod. In the sternum, intact bone, wire fixation with exact apposition, and with a gap were identified separately. Although further studies are needed to assess the accuracy of the method to detect different levels of diastases, the method appears to be feasible.
Keyphrases
  • bone mineral density
  • high frequency
  • multiple sclerosis
  • soft tissue
  • body composition
  • bone loss
  • case control