Experimental validation of a FLUKA Monte Carlo simulation for carbon-ion radiotherapy monitoring via secondary ion tracking.
Pamela Ochoa-ParraLuisa SchweinsNelly AbbaniLaura Ghesquière-DiérickxTim GehrkeJan JakubekLukas MarekCarlos GranjaFabian DinkelGernot EchnerMarcus WinterAndrea MairaniSemi HarrabiOliver JäkelJürgen DebusMária MartišíkováLaurent KelleterPublished in: Medical physics (2024)
Our study confirms the potential of our simulation framework to investigate the performance of monitoring inter-fractional anatomical changes in patients undergoing CIRT using secondary nuclear charged fragments escaping from the irradiated patient.