Comparison of Volume Measurements and Bioimpedance Spectroscopy Using A Stand-on Device for Assessment of Unilateral Breast Cancer-Related Lymphedema.

Leigh C WardBelinda M ThompsonLouise A KoelmeyerLouise A Koelmeyer
Published in: European journal of breast health (2024)
BIS was superior for identifying BCRL compared with volume measurements, reaffirming the value of this technique. However, it is recommended that BIS be used in conjunction with comprehensive evaluation of symptoms and clinical presentation. The proposed graphical method provides a simple and easily interpretable approach to compare and define concordance between the two commonly used methods for BCRL assessment namely limb volume and BIS L-Dex indices. The existing BIS (L-Dex) thresholds for presence of BCRL were also validated.