Study of the Dependency on Magnetic Field and Bias Voltage of an AC-Biased TES Microcalorimeter.

L GottardiJ AdamsC BaileyS BandlerM BruijnJ ChervenakM EckartF FinkbeinerR den HartogH HoeversR KelleyC KilbourneP de KorteJ van der KuurM LindemanF PorterJ SadlierS Smith
Published in: Journal of low temperature physics (2012)
At SRON we are studying the performance of a Goddard Space Flight Center single pixel TES microcalorimeter operated in an AC bias configuration. For x-ray photons at 6 keV the pixel shows an x-ray energy resolution ΔEFWHM =3.7 eV, which is about a factor 2 worse than the energy resolution observed in an identical DC-biased pixel. In order to better understand the reasons for this discrepancy we characterised the detector as a function of temperature, bias working point and applied perpendicular magnetic field. A strong periodic dependency of the detector noise on the TES AC bias voltage is measured. We discuss the results in the framework of the recently observed weak-link behaviour of a TES microcalorimeter.