Experimental assessment of the performance of vitreous cutters with fluids with different rheological properties.

Irene NepitaRodolfo RepettoAndrea DoderoSilvia ViciniMariantonia FerraraMario R RomanoAlessandro Stocchino
Published in: Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie (2021)
All tested vitreous probes are very sensitive to changes in fluid rheology. SB cutters produce smaller flow rates compared with DB ones of the same caliber; however, they are less sensitive to fluid properties at low aspiration pressures. The use of vitreous substitutes for test performance guarantees comparability between flow rate results achieved with different vitrectomy systems operating in different media. This outcome is further confirmed by the low values of estimated flow rate relative errors.