Login / Signup

Sensitivity and specificity of swept source optical coherence tomography (SS-OCT) for detecting macular pathologies versus spectral domain (SD-OCT).

Elizabeth YeuJohn P BerdahlPreeya K GuptaMichael Patterson
Published in: Journal of cataract and refractive surgery (2024)
Using retinal SS-OCT biometer scans as a replacement of the dedicated macular SD-OCT for screening or diagnosing macular health would not be appropriate due to its low sensitivity. SS-OCT biometer may potentially fail to identify approximately one-fourth of patients who actually have the disease. Therefore, the final decision on macular health should be based on the gold standard SD-OCT scans. When full macular SD-OCT scans are not accessible, the limited retinal scan information from SS-OCT biometer may still provide useful insights into the macular health.
Keyphrases