Impact of sweep gas flow on extracorporeal CO2 removal (ECCO2R).

Stephan StrassmannMichaela MertenSimone SchäferJonas de MollDaniel BrodieAnders LarssonWolfram WindischChristian Karagiannidis
Published in: Intensive care medicine experimental (2019)
The influence of sweep gas flow on the CO2 removal capacity of ECCO2R systems depends predominantly on blood flow rate and membrane lung surface area. In this model, considerable CO2 removal occurred only with the larger membrane lung surface of 0.8m2 and when blood flow rates of ≥ 900 mL/min were used.