Optimizing Precision: A Trajectory Tract Reference Approach to Minimize Complications in CT-Guided Transthoracic Core Biopsy.
Stella Chin-Shaw TsaiTzu-Chin WuFrank Cheau-Feng LinPublished in: Diagnostics (Basel, Switzerland) (2024)
The advent of computed tomography (CT)-guided transthoracic needle biopsy has significantly advanced the diagnosis of lung lesions, offering a minimally invasive approach to obtaining tissue samples. However, the technique is not without risks, including pneumothorax and hemorrhage, and it demands high precision to ensure diagnostic accuracy while minimizing complications. This study introduces the Laser Angle Guide Assembly (LAGA), a novel device designed to enhance the accuracy and safety of CT-guided lung biopsies. We retrospectively analyzed 322 CT-guided lung biopsy cases performed with LAGA at a single center over seven years, aiming to evaluate its effectiveness in improving diagnostic yield and reducing procedural risks. The study achieved a diagnostic success rate of 94.3%, with a significant reduction in the need for multiple needle passes, demonstrating a majority of biopsies successfully completed with a single pass. The incidence of pneumothorax stood at 11.1%, which is markedly lower than the reported averages, and only 0.3% of cases necessitated chest tube placement, underscoring the safety benefits of the LAGA system. These findings underscore the potential of LAGA to revolutionize CT-guided lung biopsies by enhancing procedural precision and safety, making it a valuable addition to the diagnostic arsenal against pulmonary lesions.