Effect of Pulsatility on Shear-induced Extensional Behavior of Von Willebrand Factor.
Yi WangKhanh T NguyenEsraa IsmailLeslie DonoghueGuruprasad A GiridharanPalaniappan SethuXuanhong ChengPublished in: Artificial organs (2021)
Pulsatile flow reduces unfolding of vWF compared to continuous flow and a lower pulse frequency of 20 pulses/minute yielded comparable vWF unfolding to 60 pulses/minute. These findings could shed light on non-surgical bleeding associated with the loss of pulsatility following CF-VAD placement.