Evaluation of novel radiation protection devices during radiologically guided interventions.

Maria E V LarssonPernilla I JonassonPetra S ApellPeter P KearneyCharlotta J Lundh
Published in: CVIR endovascular (2024)
A ceiling mounted lead glass shield is an effective radiation protection for the head, but in clinical practice, optimal positioning of a ceiling mounted lead shield may not always be possible, particularly during complex cases when radiation protection may be most relevant. Added protection using these novel guards may compliment the shielding effect of the ceiling mounted lead shield. The head protector stand-alone did not provide sufficient protection of the head. The extended thyroid protector stand-alone provided sufficient protection in the lower and middle sections of the head and neck.
Keyphrases
  • clinical practice
  • optic nerve
  • physical activity
  • radiation induced
  • radiation therapy