Reducing cardiac implantable electronic device-induced artefacts in cardiac magnetic resonance imaging.

Aino-Maija VuorinenLauri LehmonenJarkko KarvonenMiia HolmströmSari KivistöTouko Kaasalainen
Published in: European radiology (2022)
• Cardiac implantable electronic device (CIED)-induced metal artefacts may significantly diminish the diagnostic value of an MRI, particularly in cardiac MRIs. • Raising the ipsilateral arm relative to the CIED generator is a cost-free, straightforward method to significantly reduce CIED-induced artefacts on cardiac MRIs in patients with a left-sided generator. • Right-sided generator implantation reduces artefacts compared with left-sided implantation and could be considered in CIED patients requiring subsequent cardiac MRIs to ensure adequate image quality in the future.