A Deep-Learning-Based Guidewire Compliant Control Method for the Endovascular Surgery Robot.
Chuqiao LyuShuxiang GuoWei ZhouYonggan YanChenguang YangYue WangFanxu MengPublished in: Micromachines (2022)
Endovascular surgery is a high-risk operation with limited vision and intractable guidewires. At present, endovascular surgery robot (ESR) systems based on force feedback liberates surgeons' operation skills, but it lacks the ability to combine force perception with vision. In this study, a deep learning-based guidewire-compliant control method (GCCM) is proposed, which guides the robot to avoid surgical risks and improve the efficiency of guidewire operation. First, a deep learning-based model called GCCM-net is built to identify whether the guidewire tip collides with the vascular wall in real time. The experimental results in a vascular phantom show that the best accuracy of GCCM-net is 94.86 ± 0.31%. Second, a real-time operational risk classification method named GCCM-strategy is proposed. When the surgical risks occur, the GCCM-strategy uses the result of GCCM-net as damping and decreases the robot's running speed through virtual resistance. Compared with force sensors, the robot with GCCM-strategy can alleviate the problem of force position asynchrony caused by the long and soft guidewires in real-time. Experiments run by five guidewire operators show that the GCCM-strategy can reduce the average operating force by 44.0% and shorten the average operating time by 24.6%; therefore the combination of vision and force based on deep learning plays a positive role in improving the operation efficiency in ESR.
Keyphrases
- deep learning
- single molecule
- minimally invasive
- artificial intelligence
- convolutional neural network
- coronary artery bypass
- machine learning
- surgical site infection
- human health
- magnetic resonance imaging
- quality improvement
- magnetic resonance
- risk assessment
- acute coronary syndrome
- coronary artery disease
- high intensity
- atrial fibrillation
- aortic dissection