Operation and Performance of a Mass-Selective Cryogenic Linear Ion Trap.
Larry F TeslerAdam P CismesiaMatthew R BellLaura S BaileyNicolas C PolferPublished in: Journal of the American Society for Mass Spectrometry (2018)
We report on the performance of a cryogenic 2D linear ion trap (cryoLIT) that is shown to be mass-selective in the temperature range of 17-295 K. As the cryoLIT is cooled, the ejection voltages during the mass instability scan decrease, which results in an effective mass shift to lower m/z relative to room temperature. This is attributed to a decrease in trap radius caused by thermal contraction. Additionally, the cryoLIT generates reproducible mass spectra from day-to-day, and is capable of performing stored waveform inverse Fourier transform (SWIFT) mass isolation of fragile N2-tagged ions for the purpose of background-free infrared dissociation spectroscopy. Graphical Abstract ᅟ.