Displacement of scan body during screw tightening: A comparative in vitro study.

JungHan KimKeunbada SonKyu-Bok Lee
Published in: The journal of advanced prosthodontics (2020)
The horizontal displacement in all groups was less than 10 µm. With the hand tightening torque, a high vertical displacement of over 100 µm occurred in PEEK scan bodies (Myfit and Dentium). Therefore, it is recommended to apply a tightening torque of 5 Ncm instead of a hand tightening torque.