Complex transcatheter left atrial appendage closure using a tailored trans-jugular approach simulated by 3D printing: a case report.
Vlad CiobotaruVictor-Xavier TadrosClaire A MartinSebastien HascoetPublished in: European heart journal. Case reports (2022)
This is the first-reported clinical case of a complex transcatheter LAAc through a trans-jugular approach. Simulating the patient's anatomy with a laser sintering 3D-printed model showed why the transfemoral approach failed, validated the trans-jugular procedure, enabled selection of the simple curve access sheath that had the most direct trajectory towards the LAA, confirmed that transseptal puncture was possible, allowed determination of the angle of puncture, enabled selection of the most appropriate LAA device and had a very low cost compared with planning software or other printing methods.