Occurrence and predictors of acute stent recoil-A comparison between the xience prime cobalt chromium stent and the promus premier platinum chromium stent.
Rutger J van BommelMiguel E LemmertNicolas M van MieghemRobert-Jan van GeunsRon T van DomburgJoost DaemenPublished in: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (2017)
No significant differences were observed in the extent of acute stent recoil between the Xience Prime and the Promus Premier stent. A larger stent/vessel ratio, a larger balloon/stent ratio, and direct stenting were associated with significant acute stent recoil ≥10%. © 2017 Wiley Periodicals, Inc.