Login / Signup

Impact of errors in spot size and spot position in robustly optimized pencil beam scanning proton-based stereotactic body radiation therapy (SBRT) lung plans.

Suresh RanaAnatoly B Rosenfeld
Published in: Journal of applied clinical medical physics (2021)
The increasing spot sizes resulted in decreased target coverage and dose homogeneity. Similarly, the decreasing spot sizes led to a loss of target coverage, overdosage, and degradation of dose homogeneity. The addition of spot size and position errors to plan robustness parameters (setup and range uncertainties) increased the target coverage loss and decreased the dose homogeneity.
Keyphrases
  • radiation therapy
  • resting state
  • affordable care act
  • patient safety
  • healthcare
  • health insurance
  • adverse drug
  • emergency department
  • squamous cell carcinoma
  • electronic health record
  • drug induced