A quantitative analysis of optimum design for rigid ankle foot orthoses: The effect of thickness and reinforcement design on stiffness.

Panagiotis E ChatzistergosNicola EddisonEvangelia Ganniari-PapageorgiouNachiappan Chockalingam
Published in: Prosthetics and orthotics international (2023)
For a specific AFO geometry and loading magnitude, there is a thickness threshold below which the AFO cannot effectively resist flexion and buckles. FE modelling showed that stiffness is maximised when reinforcements are placed at the anterior-most position possible. This key finding was also experimentally confirmed. The stiffness of an AFO reinforced according to standard practice with lateral and medial ribbing was 4.4 ± 0.1 Nm/degree. Instructing the orthotic technician to move the ribbings anteriorly increased stiffness by 22%. Further stiffening is achieved by ensuring the reinforcements extend from the footplate to at least two-thirds of the AFO's total height.