MLC parameters from static fields to VMAT plans: an evaluation in a RT-dedicated MC environment (PRIMO).
Lucia PaganiniGiacomo ReggioriAntonella StravatoValentina PalumboPietro MancosuFrancesca LobefaloAnna GaudinoAntonella FogliataMarta ScorsettiStefano TomatisPublished in: Radiation oncology (London, England) (2019)
MC simulations in the PRIMO environment were in agreement with measurements for the HD-120 MLC in a 10 MV FFF beam from a Varian EDGE linac. This result allowed to consistently compare clinical cases, showing the possible use of PRIMO as an independent dose calculation check tool.