Cognitive performance and brain dynamics during walking with a novel bionic foot: A pilot study.

Romain MeeusenPierre CherelleBruno TassignonJeroen Van CutsemBart RoelandsFelipe Gomez MarulandaDirk J LefeberBram VanderborghtRomain Meeusen
Published in: PloS one (2019)
In contrast to transtibial amputees, transfemoral amputees required more attentional demands during walking with Ankle Mimicking Prosthetic foot 4.0 compared to the current passive prosthetic device and able-bodied individuals (reaction time and accuracy: p≤0.028). Since risk of falling is associated with altered attentional demands, propulsive forces of the novel device need to be better controlled for transfemoral amputees. No motor-related cortical potentials at Cz were observed in transfemoral amputees walking with the novel prosthesis, whereas motor-related cortical potentials between transtibial amputees and able-bodied individuals during walking at normal speed did not differ. The first positive electro-physiological peak deflection appeared during toe-off phase and showed higher activity within the underlying brain sources in transtibial amputees walking with Ankle Mimicking Prosthetic foot 4.0 compared to able-bodied individuals. The required higher neural input to accomplish the same physical activity compared to able-bodied individuals is possibly due to the limited acclimation period to the novel device and consequently increased afferent sensory feedback for postural control.