Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures.
Christopher ThomasJohn B WelshSpencer LuJ Michael GrayPublished in: Journal of diabetes science and technology (2020)
Wearable components of the G6 CGM System retain basic functionality and data integrity after exposure to simulated therapeutic and diagnostic radiologic procedures. The devices are unlikely to be affected by x-irradiation used in typical imaging studies. Simulated MRI procedures create displacement force, minimal heating, and current in sensor/transmitter pairs.