Inverse association between blood pressure and pulse oximetry accuracy: an observational study in patients with suspected or confirmed COVID-19 infection.
Colin J CrooksJoe WestJo MorlingMark SimmondsIrene JuurlinkSteve BriggsSimon CruickshankSusan Hammond-PearsDominick ShawTim CardAndrew W FogartyPublished in: Emergency medicine journal : EMJ (2022)
Care needs to be taken in interpreting pulse oximetry measurements in patients with lower systolic and diastolic BPs, and HRs, as oxygen saturation is overestimated as BP and HR decrease. Confirmation of the oxygen saturation with an ABG may be appropriate in some clinical scenarios.