Single-Center Experience on the Elective Hybrid Combination of Single Perclose + Angio-Seal VIP 8F Compared With Standard Dual Perclose During Percutaneous Endovascular Aortic Aneurysm Repair.
Paolo SpathFilippo MaioliFederica CampanaTeresa GabelliniAnnalisa PerulliMichele LeoneFrancesco GiacchiGiacomo Di IasioMassimiliano MariniSilvia MassiniSara PomattoCecilia AngheràSalvatore TarantiniPublished in: Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists (2024)
This study introduces a novel elective hybrid technique combining a single Perclose device with a single Angio-Seal VIP 8Fr for percutaneous endovascular abdominal aortic interventions. Results for hybrid technique showed 100% technical success and efficient immediate hemostasis, while costing less than standard dual Perclose procedure. Both procedures did not require surgical conversions. Despite being a single-center study, it demonstrates potential benefits of the intentional application of this hybrid technique towards minimally invasiveness. Obesity and dual Perclose technique were identified as independent risk factors for hemostasis failure, reaffirming the hybrid technique procedure's efficacy as well as and non-inferiority to standard procedure.