A transient flow reactor for rapid gas switching at atmospheric pressure.
Eric A HighEsther LeeChristian ReecePublished in: The Review of scientific instruments (2023)
Herein, we present a design for a transient flow reactor system with high detection sensitivity and minimal dead volume, such that it is capable of sub-second switching of the gas stream flowing through a catalytic bed. We demonstrate the reactor's capabilities for step transient, pulse, and stream oscillation experiments using the model system of CO oxidation over Pd catalysts, and we find that we are able to precisely model step transients for CO oxidation using a pseudo-homogenous-packed bed reactor model. The design principles leading to minimal gas hold-up time and increased sensitivity that are described in this paper can be implemented into existing flow reactor designs with minimal cost, providing a readily accessible alternative to the existing transient instrumentation.