Calibration method of crosshair light sharing PET detector with TOF and DOI capabilities.
Eiji YoshidaFujino ObataTaiga YamayaPublished in: Biomedical physics & engineering express (2023)
By applying the developed calibration method, the CLS PET detector achieves the CRT of 251 ps and the DOI resolution of 3.3 mm.