A setup for hard x-ray time-resolved resonant inelastic x-ray scattering at SwissFEL.
Hui-Yuan ChenRolf B VersteegRoman MankowskyMichele PuppinLudmila M D LeroyMathias SanderYunpei DengRoland Alexander OggenfussThierry ZamofingPirmin BöhlerClaude PradervandAldo MozzanicaSeraphin VetterGrigory SmolentsevLinda KerkhoffHenrik Till LemkeMajed CherguiGiulia F ManciniPublished in: Structural dynamics (Melville, N.Y.) (2024)
We present a new setup for resonant inelastic hard x-ray scattering at the Bernina beamline of SwissFEL with energy, momentum, and temporal resolution. The compact R = 0.5 m Johann-type spectrometer can be equipped with up to three crystal analyzers and allows efficient collection of RIXS spectra. Optical pumping for time-resolved studies can be realized with a broad span of optical wavelengths. We demonstrate the performance of the setup at an overall ∼180 meV resolution in a study of ground-state and photoexcited (at 400 nm) honeycomb 5d iridate α-Li 2 IrO 3 . Steady-state RIXS spectra at the iridium L 3 -edge (11.214 keV) have been collected and are in very good agreement with data collected at synchrotrons. The time-resolved RIXS transients exhibit changes in the energy loss region <2 eV, whose features mostly result from the hopping nature of 5d electrons in the honeycomb lattice. These changes are ascribed to modulations of the Ir-to-Ir inter-site transition scattering efficiency, which we associate to a transient screening of the on-site Coulomb interaction.