Reduced fracture load of dental implants after implantoplasty with different instrumentation sequences. An in vitro study.
Isabelle C JorioBogna StawarczykThomas AttinPatrick R SchmidlinSahrmann PhilippPublished in: Clinical oral implants research (2021)
Both implantoplasty and cyclic loading were shown to reduce the implants' maximum bending strength. Cyclic loading in a laboratory masticator, simulating a five-year equivalent of chewing, did not result in fractured implants. Since neighbouring tooth injury was assessed often, care should be taken with the selection of suitable instruments.