Locking Multi-Laser Frequencies to a Precision Wavelength Meter: Application to Cold Atoms.
Junwoo KimKeumhyun KimDowon LeeYongha ShinSungsam KangJung-Ryul KimYoungwoon ChoiKyungwon AnMoonjoo LeePublished in: Sensors (Basel, Switzerland) (2021)
We herein report a simultaneous frequency stabilization of two 780-nm external cavity diode lasers using a precision wavelength meter (WLM). The laser lock performance is characterized by the Allan deviation measurement in which we find σy=10-12 at an averaging time of 1000 s. We also obtain spectral profiles through a heterodyne spectroscopy, identifying the contribution of white and flicker noises to the laser linewidth. The frequency drift of the WLM is measured to be about 2.0(4) MHz over 36 h. Utilizing the two lasers as a cooling and repumping field, we demonstrate a magneto-optical trap of 87Rb atoms near a high-finesse optical cavity. Our laser stabilization technique operates at broad wavelength range without a radio frequency element.