Comparison of rhinomanometric and computational fluid dynamic assessment of nasal resistance with respect to measurement accuracy.

Nora SchmidtHans BehrbohmLeonid GoubergritsThomas HildebrandtJan Brüning
Published in: International journal of computer assisted radiology and surgery (2022)
Biases introduced during RMM measurements, by either the chosen device, the operator or other aspects as for example the nasal cycle, are not strong enough to explain the gross differences commonly reported between RMM- and CFD-based measurement of nasal resistance. Additionally, tissue movement during breathing is most likely also no sufficient explanation for these differences.
Keyphrases
  • chronic rhinosinusitis