Chiral separation of calcium channel antagonists by SFC and HPLC using different immobilized chiral stationary phases.

Rupak RajaSyed Dilshad AlamDipanjan MukhopadhyayVikas ShirsathArvind K JainZeid Abdullah ALOthmanMarcello LocatelliImran Ali
Published in: Chirality (2022)
Supercritical fluid chromatography and high-performance liquid chromatography techniques are popular for the chiral separations of different drugs and pharmaceuticals. Therefore, this article describes a comparative study of the chiral separation of some calcium channel antagonists such as verapamil, gallopamil, and nisoldipine. The columns used were Chiralpak IG and Chiralpak ID (250 mm × 4.6 mm, 5.0 μm). The separation was achieved by using a variety of mobile phases in both techniques. The retention, separation, and resolution factors in SFC were in the range of 1.36-7.30, 1.09-1.72, and 1.16-3.47, while these values in the case of HPLC were 1.03-2.42, 1.12-1.35, and 0.49-2.46. The complete resolution of gallopamil and verapamil was achieved successfully. The chiral recognition was controlled by hydrogen bondings, π-π interactions, dipole induced dipole interactions, van der Waal forces, and steric effects. SFC was found to be a better technique than HPLC because of quick separation, good separation power, economic, environment-friendly, and green technology.